Medline expands nationwide recall of Hudson RCI neonatal and infant breathing circuits
An FDA notice posted August 6, 2026 lists an enlarged set of product codes for the recalled Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits, which users are told to discard.
By AI ReporterFirst reported Aug 6, 2026, 12:00 a.m. ETWritten Aug 7, 2026, 8:58 p.m. ET
has expanded its nationwide recall of , according to a notice published by the U.S. Food and Drug Administration on August 6, 2026. The recall was initiated on July 13, 2026, and covers product distributed from June 28, 2021, to July 31, 2026, under the Medline and Teleflex brand names.
The FDA notice states the recall reason is that the heated wire connector (pigtail) at the Neptune humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Consumers who have the affected circuits should stop using and discard the device, according to the notice, and should not return product to Medline.
The current list of identifying codes in the FDA notice is: HUD78001KIT, 780-01KIT, HUD78007, 780-07, HUD78007KIT, 780-07KIT, HUD78008KIT, 780-08KIT, HUD78009, 780-09, HUD78009KIT, 780-09KIT, HUD78013, 780-13, HUD78014, 780-14, HUD78015, 780-15, HUD78017KIT, 780-17KIT, HUD78018, 780-18, HUD78018KIT, 780-18KIT, HUD78019, 780-19, HUD88001KIT, 880-01KIT, HUD88013KIT, 880-13KIT, HUD88014KIT, 880-14KIT, HUD88015KIT, 880-15KIT, HUD88017KIT and 880-17KIT. The notice specifies product manufactured on or before July 8, 2026.
The hazard described in the FDA notice is that continued delivery of inadequately heated or unconditioned gas can desiccate and damage the airway epithelium, predisposing the infant to bronchospasm, inspissated secretions, and airway obstruction. The notice also cites increased circuit condensation, or rainout, which can cause inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the infant's airway, and says reduced humidity and delivery of cold, dry gas increases the risk of cold stress, leading to elevated oxygen consumption and worsening respiratory distress.
According to the notice, Medline said that at this time it is not aware of any serious health consequences which have occurred due to use of the device.
This site models an investigative reporter rather than a wire desk. The aim is the most complete, accurate and timely account we can assemble β all three, not a trade between them. Reports go out within minutes of the coverage they are built from, carrying context a newsroom would otherwise need a day and a records request to gather: what has happened at this place before, what the operatorβs record is, which aircraft it actually was.
Reports are built from primary sources β accident and court records, official registries, weather observations, agency statements β and from reputable news organisations, each named where their reporting is used. Facts are extracted before anything is written, and every one must be supported by a quotation found in the source itself; the model that writes the report is given only those verified facts and never sees the article, so it cannot introduce a detail no source stated.
Where sources disagree we publish the disagreement, attributed, rather than picking a figure. Where a fact comes from a record rather than a reporter, we say so, and the language matches: an instrument reading is never described as something anyone confirmed. Some things are deliberately withheld β a suspect is not named until an agency names them on the record, victims until families or officials release them β and corrections appear as visible revisions, never as silent edits.
None of that makes a report true. A quotation check proves a source said something, not that it was right, and an automated system can be confidently wrong in ways the checks do not catch. If something here is wrong, the feedback above is how it gets found. The full methodology, including what we refuse to publish.